Applying Previous Volume Sensor Calibration Result

image_pdfimage_print

The calibration history shows previous volume sensor calibration results, which include weights for the Sample (Left) Scale and Buffer (Right) Scale. If you want to reload previously calibrated volume sensors due to errors in the current calibration, these calibrations will be saved. Follow the instructions below to load a previous calibration result.

 

To apply the calibration result:

  1. On the CALIBRATION WIZARD tab in the SYSTEM SETTINGS menu, then tap HISTORY.
  2. Select the specific calibration result in the calibration history list and tap USE SELECTED VALUE to apply that calibration result to your current µPULSE hardware.

    Tip: Find older calibration results by selecting a specific date.

  1. Calibration values have successfully been updated! Tap the CHECK SCALES button to check the current scales.
Loading Previous Sensor Calibration
Loading Previous Sensor Calibration

Important: If you encounter any issues loading the previous calibration from the history, contact support@formulatrix.com for assistance.