Setting Up Auto Profile
The µPULSE allows you to create a profile consisting of desired experiment settings for concentrating sample or performing buffer exchange. The system tracks volume as it runs sequences based on the values you have entered when designing a profile.
To set up a new profile, tap AUTO PROFILE on the Home Screen. Then choose between these 3 options:
- Tap the CONCENTRATE tab if you want to concentrate a sample 50 mL or less.
- Tap the CONCENTRATE > 50 ML tab if you want to concentrate more than 50 mL of sample.
- Tap the BUFFER EXCHANGE tab if you want to exchange buffer solution for another from the buffer (right) tube. The reagent type can be the same or different as the carrier reagent.
Tip: Tap the Help button to get acquainted with the software and learn about what you can do in the AUTO PROFILE menu.
Creating a New Profile
- Tap AUTO PROFILE on the Home Screen and select one of the options:
To create and run a concentrate profile, follow the instructions below:
- From the CONCENTRATE tab, enter the Starting Volume and the Starting Concentration, then the Final Volume OR the Final Concentration.
Note: The final concentration or final volume will be calculated based on the other inputs.
- Select the Membrane Cut-Off Size. Then, continue to step 2.
- Tap the CONCENTRATE > 50 ML tab.
- Fill in the parameters for Starting Volume in Tube 01 and 02, then the Final Volume.
- Then, continue to step 2.
- Select the BUFFER EXCHANGE tab. Input the Starting Volume and Final Volume. The rest of the parameters will adjust automatically.
- Click on the exchange volume calculator and select the extent of the buffer exchange by inputting the percentage required. It will calculate the volume of the Buffer B required for the exchange.
- The step size indicates the volume of buffer B exchanged at each step. You can reduce this volume according to the requirement.
- Continue to step 2 to load the saved profile for buffer exchange.
- Click the ADVANCED SETUP EXPLORER to change the pressure settings.
- Click EDIT to change settings.
-
-
- In the General Setup tab, the Safety Threshold can be adjusted. You can also choose the priming option between BUFFER or SAMPLE and define the Priming Volume.
- Click the CYCLE SETUP tab to to adjust the diaphragm pressure, return valve pressure, and fill/empty time.
-
Note: Tap the Edit button to adjust Return Valve Pressure.
- Tap SAVE SETUP AS to save the customized settings.
- Click LOAD SETUP to use the saved settings.